Removing The Shock Wave Head - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

LITHOSTAR MODULARIS
1
2
V
V
Fig. 3

Removing the shock wave head

System con-
1 LITHOSTAR MODULARIS
figuration
and
n = several
1 SIREMOBIL Iso-C
To support arm S0050
Adjustment of iso-center on the shock
wave head; refer to chapter "Iso-cen-
ter with X-ray - to support arm serial
Replace
number 0050".
shock wave
From support arm S0051
head
Adjustment of iso-center on the
SIREMOBIL Iso-C; refer to chapter
"Iso-center with X-ray - from support
arm serial number 0051".
Remove the covers of the shock wave head (refer to "Shock wave head covers").
Remove the coupling bellows, carefully pulling the small pin out of the air suction hose
inside the coupling bellows (refer also to Fig. 8 and Fig. 9).
Loosen the four screws (V/Fig. 3) on the iso-center phantom using an Allen key with a
guide pin, but do not remove them.
Flip the iso-center phantom toward the front and completely remove it (this may be
difficult).
- If the iso-center phantom has a serial number (2/Fig. 3) it can be removed by using two
screw pulleys (1/Fig. 3) on the shock wave head.
Siemens AG
Medical Solutions
1
Fig. 4
1 LITHOSTAR MODULARIS
and
n SIREMOBIL Iso-C
It is important to follow this workflow to
avoid having to readjust the iso-center on all
SIREMOBILE!
Prior to replacing the shock wave head, view
the iso-center (the three C-arm positions) and
if possible, save it or make a hardcopy.
Replace shock wave head.
Check iso-center; the setting must be
identical to that of the defective shock wave
head.
If an adjustment is necessary, it must be made
via the shock wave head setting; refer to
chapter "Iso-center with X-ray - to support
arm serial number 0050".
SPL1-130.840.02
Rev. 03
04.05
U
Page 3 of 34
MODULARIS Uro Plus
CS PS 24
4 - 3
4