Software Download With Windows 95 / 98 / Nt - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Service software

Software download with Windows 95 / 98 / NT

Do not select "capture text file" when downloading the software.
German Windows:
In the "Übertragung" menu, select ... Datei senden
In the "Datei senden" window: Dateiname: click
"Durchsuchen"; select the file with the "bin" extension
on the diskette inserted (Laufwerk beachten)
Protokoll: select Xmodem . Click Senden
No
During the data transfer, "1" is displayed on board D3 and the decimal point flashes.
After the transfer, a test is executed. ("2" is displayed on board D3; "0" or "7" when complete.)
Press the "n" key and
enter the name
Check the system-specific settings (language, ECG triggering, pulse rate and table)
Siemens AG
Medical Solutions
Log on, as described on page 3-2 or 3-3
Insert the diskette containing the software into drive A
Press the "s" key and then press "1"
Disregard the error messages
Switch the system off / on
Press the "h" key to display the text
Is correct software
version displayed?
Was software
Yes
transferred to a new
D3 board?
No
Set the date and time. Check Ultrasound default values.
For additional procedures, refer to pages 3-2 or 3-3
SPL1-130.840.02
Rev. 03
English Windows:
In the "Transfer" menu, select ... Send File
In the "Send File" window: File name: click
"Browse"; select the file with the "bin" extension on
the diskette inserted. (observe drive)
Protocol: select Xmodem . Click Send
Page 13 of 18
04.05
CS PS 24
3 - 13
3
MODULARIS Uro Plus

Advertisement

Table of Contents
loading

Table of Contents