Tools And Measurement Devices Required; Auxiliary Devices And Documents Required - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

1 - 2
Initial measured value
The equivalent leakage current measurement was performed at the factory and the value
measured was entered in test protocol 1. The measurement was made at the line voltage
and line frequency indicated in test protocol 1.
If the line voltage or the line frequency on-site deviates from the information indicated
upon delivery of the Modularis Uro Plus, the values given are invalid. The values should
be marked invalid (crossed out with the comment "invalid values" and the service engi-
neer should sign and date this copy).
In this case, the equivalent leakage current must be measured again. The value may not
exceed 1 mA according to DIN VDE 0751, Part 1.
The initial value measured must be documented.
Repeat measurements
When service or repair work is performed on the primary power supply circuit (e.g. repairs
to the power-on circuit), the equivalent leakage current test must be repeated. The values
measured in the subsequent test may not exceed the threshold value of 1 mA as specified
in VDE 0751, Part 1. In addition, they may not exceed the initial measured value by more
than 50%. If the value exceeds this threshold, the system must be repaired. The value
measured must be documented.

Tools and measurement devices required

NOTE
Standard service equipment
Service PC (see Intranet Service Laptop for CSE's
PC connection cable, 5 m
ESD equipment
Protective conductor meter
Digital multimeter, e.g. FLUKE 187
Shock wave pressure test unit
Adapter for C-head

Auxiliary devices and documents required

SIREMOBIL Iso-C Service Instructions
Service Instructions Shock Wave Pressure Control
MODULARIS Uro Plus
The specified articles are listed in the SCT (Catalog Service Tools)
if nothing else is indicated (the STC is a component of the Spare
Parts Catalog).
SPL1-130.840.02
Rev. 03
04.05
General
99 00 440 RE999
44 15 899 RV090
99 94 831
30 95 408 J1008
98 17 347 J1008
SPR2-230.061.01..
SPL2-120.074.01..
Page 2 of 2
CS PS 24
Medical Solutions
1
1
Siemens AG