Siemens MODULARIS Uro Plus Service Instructions Manual page 92

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

6 - 10
A
Fig. 16
Set all slide controls all the way to the left.
The image flip
The corresponding ultrasound probe is selected during switch-on. If there is a second
ultrasound probe, the select the probe with the
Check whether the ultrasound arm is as shown in Fig. 15; if not, position it accordingly.
The long slot must be visible (1/Fig. 15).
Checking the distance: Height (16a/Fig. 16)
Press the
Move the track ball until the corss "+1" is in the center of the isocenter crosshair.
The value displayed for "D1" (G/Fig. 16) must be between 33 and 36 mm; if not, move the
probe appropriately in the longitudinal axis.
Press the
Move the cross "+2" using the track ball onto the longitudinal axis of the target line in the
isocenter (M/Fig. 16a).
Press the
Move the cross using the track ball horizontally into the center of the white area
(16a/Fig. 16) of the isocenter phantom
The value displayed for "D2" (H/Fig. 16) must be between 1 and 5 mm. This is only valid
for the default values, see instructions, SPL1-130.038.01...
Press the
MODULARIS Uro Plus
G
H
E
key may not be selected.
key.
key.
key.
key.
SPL1-130.840.02
Rev. 03
Iso-center with ultrasound
16a
M
B
2
D
C
C
key.
Page 10 of 16
04.05
CS PS 24
Image detail sche-
:
matic
16b
C
2
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

Table of Contents