6 - 4
Fig. 7
Make the following basic settings on the ultrasound unit:
Value:
5
(Fig. 7)
35 dB (Fig. 7)
Set all slide controls all the wayto the left.
The image flip key
The corresponding ultrasound probe is selected on switching on. If a second ultrasound
probe is present, then select the probe with the
Check that the ultrasound arm is as shown in Fig. 1, if not align it.
The longitudinal groove must be visible (1/Fig. 8).
MODULARIS Uro Plus
Setting with:
Set trackball to Edge
FV
(white background)
then press
and
set value with
"Select" controller.
Set trackball to DR
DR
(white background)
then press
and
set value with
"Select" controller.
must not be selected.
SPL1-130.840.02
Rev. 03
04.05
Iso-center with ultrasound
Fig. 8
German
English
Edge
Côté
Dyn Rng
DR
key.
Page 4 of 16