Select Option: C - Siemens MODULARIS Uro Plus Service Instructions Manual

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Service software

select option: c

Press the "c" key to display the following text::
=> C
Card Data:
# Serial number Type -------< units >--------
GOLD init-cnt
PPU
------------------------------------------------------------------------------
1:
2:
3:53148300000000
4:53149100000001
5:53148300000000
6:53149100000001
7:
8:
9:
10:53149100000001
-- end of card data --
Text:
Explan.:
- *1: sequential number of the card inserted.
- *2: serial number of the card.
- *3: a "gold" card was inserted, i.e. the input slot is closed.
- *4: init-counter in the above example is init=2, i.e. the gold card was logged onto two
different D3 boards.
- *5: n.a.
- *6: n.a.:
Card ejected
Card in system
Text:
Explan.:
- *1:sequential number of the card inserted
- *2: serial number of the card
- *3: a "pay per use" card was inserted
- *4: shock waves still present on the card
- *5: internal system card counter
- *6: number of shock waves registered on inserting the ppu card
Card ejected
Card in system
Siemens AG
Medical Solutions
...
on card
int.
G init=2
....
....
P
89956
0
G init=3
....
....
P
84956
0
5000
P
84956
5000
4:
531...0
G
init=2
*1
*2
*3 *4
- *7: Date of ejection
- *8: System shock wave counter value when card ejected
- *7: Text: "Card locked to unit"
- *8: ............
5:
531...1
P
*1
*2
*3
- *7: Date of ejection
- *8: System shock wave counter value when card ejected
- *7: Text: "Card locked to unit"
- *8: ............
SPL1-130.840.02
Rev. 03
eject-date/time
...
from card
04/13/1999 15:48.11
0
04/13/1999 16:00.21
04/13/1999 16:40.51
04/13/1999 17:45.00
0
card locked to unit
....
.....
*5
*6
89956
0
*4
*5
Page 9 of 18
04.05
CS PS 24
pulse-count
system
418
418
418
418
......
04/13 ...
418
*7
*8
0
04/13 ... 418
*6
*7
*8
MODULARIS Uro Plus
3 - 9
3

Advertisement

Table of Contents
loading

Table of Contents