Iso-center with ultrasound
Checking the distance: Side (16b/Fig. 16)
•
Press the
•
Move the cross "+2" using the track ball onto the horizontal axis of the target line at the
height of isocenter (C/Fig. 16).
•
Press the
•
Move the cross using the track ball horizontally into the center of the white area of the
isocenter phantom.
•
If the center point of the white area (16b/Fig. 16) is not in the center
(permitted difference: + 0.5 mm from the center):
- Make a note of the value of D2 (with mathematical sign):
If the display of the ball (16b/Fig. 16) appears to the left of the isocenter, the correction
value must have a negative sign in front (to the right is a positive sign in front).
- Press the
- Press the "F4" key.
- Press the "5" key.
- Select "MUP" with the track ball.
- Press the
- Enter the password.
- Use the track ball to select the appropriate probe during image tilt.
German:
English:
- Press the
- Enter half the value of D2 (H/Fig. 16).
Example:(H/Fig. 16): D2 = 1,2 mm
- Press the
- Press the
- Check the difference; if white area (16b/Fig. 16) is not in the center of the isocenter,
repeat the procedure.
•
Press the
•
After adjusting the mount for the ultrasound probe, attach it on the mount (7/Fig. 1).
For each probe, check whether the adjustment in this position is also okay. Do not make
any change to the image tilt adjustment.
•
Following the check, store the parts in the appropriate transport cases.
Siemens AG
Medical Solutions
key twice (cross "+1" is not needed for this measurement).
key.
key.
key.
Bildneigung Sektor
French:
Bildneigung Array
Image Tilt Sector
Spanish:
Image Tilt Array
key.
key.
key twice.
key.
SPL1-130.840.02
Rev. 03
Basculement d'image sondes sect.
Basculement d'image sondes conv.
Incl. imagen transd. sect mecanico
Inclinar imagen transductor curvo
⇒
Enter: -06 .
Page 11 of 16
04.05
CS PS 24
6 - 11
MODULARIS Uro Plus