6 - 12
Fig. 17
Checking the target on the Sonoline Prima
NOTICE
•
Set the value (D/Fig. 17,18) to 180 using the knob
•
Slide the ultrasound probe all the way forward to the shock wave head.
•
Press the
•
Use the track ball to move to the center of the crosshairs. In the display (F/Fig. 17), a
value of - see Table - must appear that is relative to the probe.
Probe
Sector (1/Fig. 2)
Array (2/Fig. 2)
•
Press the
•
Move the ultrasound probe all the way in the other direction.
•
Press the
•
Use the track ball to move to the center of the crosshairs. In the display (F/Fig. 18), a
value of - see Table - must appear that is relative to the probe. If this is not the case, the
adjustment procedure must be repeated for the potentiometer in the ultrasound arm or
the ultrasound arm must be replaced.
Probe
Sector (1/Fig. 2)
Array (2/Fig. 2)
•
Press the
MODULARIS Uro Plus
F
D
The isocenter must be checked with every probe type and after
every replacement of a probe.
key and select the measuring point
Select using
key ("Freeze" is deselected).
key and select the measuring point
Select using
key ("Freeze" is deselected).
SPL1-130.840.02
Rev. 03
Iso-center with ultrasound
Fig. 18
.
.
Value for D1
15 ± 1 mm
with default values,
see SPL1-130.038.01
13 ± 1 mm
.
Value for D1
153 ± 1 mm
with default values,
see SPL1-130.038.01
151 ± 1 mm
Page 12 of 16
04.05
CS PS 24
F
D
Siemens AG
Medical Solutions