Fig. 25
Install the water supply (W/Fig. 20) in the shock wave head support, following the same
procedure as for removal.
Reinstall the white teflon plate (Tp/Fig. 18).
Install the temperature sensor (T/Fig.20) with the spiral spring in the shock wave head
support following the same procedure as for removal in the reverse order; when doing
this, make sure that the spring force of the spiral spring is pushing forward against the
temperature sensor (Fig. 18).
Install the shock wave head assembly (refer to the corresponding section).
Reattach all covers (refer to the corresponding section).

From support arm serial number 0051

Remove the covers of the basic system unit and the shock wave head.
Empty the coupling circuit (refer to specific paragraph).
Remove the coupling bellows.
Remove both screws (1/Fig. 21).
Pull out the lower part (2/Fig. 21).
Remove the part with the air suction hose (1/Fig. 22).
Replace the air suction hose.
Pierce the hose two times with a needle (approx. 1mm) in the same manner the old hose
was pierced.
Reassemble in the opposite order.Ensure proper position of the Allen screws during
reassembly.