4 - 20

Board D3 or Chip card reader

Notice for
- Do not use the Gold card to troubleshoot other systems!
- Changing the Gold card too many times will render it unusable.
- After each change of the Gold card to another system, the user must confirm this on
the control unit.
- The Gold card and the board each contain an internal counter; if these counters do not
have the same value, the counter values will increase.
- If the card must be changed due to service, the switch on board D3 must be set to
service prior to beginning the service work. In this case, the counter values will not
change. However, shock wave release will be blocked but the system will initialize.
- If D3/M13 is being replaced, an internal counter on the Gold card will count up.
The Gold card will be automatically rendered unusable if the Init counter in the "Chip
Card" menu exceeds "3".
- A new Gold card must be ordered from the sales department.
After receiving the new Gold card, the old card must be returned to SPL in Erlangen.
If the card is not returned, full price will be charged.

Replacing board D3

Prior to installing a new board, if possible, read out the counter status and record the
value in the operating protocol (register protocol).
Prior to removing board D3, remove the chip card (Gold card or pay-per-use) from the
reader.
- For Gold card only:
Prior to ejecting the card, remove the reader from the covers by removing the screws
(the insertion slot must be clear).
- Eject:
Set switch on board D3 to service mode.
In the service menu, select chip card and eject.
Switch the system off. Replace board D3. Prior to installing the board, check whether the
battery is connected.
If the ultrasound localization is available, assemble the D3 board on the new ultrasound
addition.
Switch the system on.
Connect the service PC and then connect to the LITHOSTAR MODULARIS. Perform a
software download (refer to chapter 3).
Perform system-specific settings:
time, language, table right or left, pulse rate, ECG triggering.
Reset switch S2 on board D3 to normal mode.
Turn system off and then on again.
MODULARIS Uro Plus
LITHOSTAR MODULARIS with the Gold card:
SPL1-130.840.02
Rev. 03
LITHOSTAR MODULARIS
Page 20 of 34
04.05
CS PS 24
4
4
4