4 - 26
8
Fig. 30
Adjusting switches S1/S2
•
Adjustment of activating lever (Fig. 31).
•
When activating pin S1 (Fig. 29), the travel movement of the vertical column is blocked.
- When activating pin S2 (Fig. 29), both movement directions of the vertical column are
blocked.
•
The screws behind the opening (7/Fig. 29) should be loosened only. Move switch until
switching function is activated and tighten screws.
Setting the cam on the LITHOSTAR MODULARIS
NOTICE
•
Remove the knob (8/Fig. 30); lift up the ball and secure the axis from shifting.
•
To support arm serial number 0050
- Remove the countersunk screw (9/Fig. 30).
- Remove all screws from the covers.
- Firmly pull the handle apart (10/Fig. 30).
MODULARIS Uro Plus
9
10
11
Only necessary if the switch on the SIREMOBIL Iso-C cannot be
adjusted.
SPL1-130.840.02
Rev. 03
LITHOSTAR MODULARIS
S1/S2
Iso-C
S1/S2
Iso-C
Fig. 31
Page 26 of 34
04.05
CS PS 24
Docking plate
1 to 0.5 mm
Docking plate
4
4
Siemens AG
Medical Solutions