Siemens MODULARIS Uro Plus Service Instructions Manual page 89

Hide thumbs Also See for MODULARIS Uro Plus:
Table of Contents

Advertisement

Iso-center with ultrasound
Fig. 10
Checking the target on the Sonoline Adara
NOTICE
Set the value (D/Fig. 10, 11) to 180 using the knob
Slide the ultrasound probe all the way forward to the shock wave head.
Press the
Use the track ball to move to the center of the crosshairs. In the display (F/Fig. 10), a
value of - see Table - must appear that is relative to the probe.
Probe
Sector (1/Fig. 2)
Array (2/Fig. 2)
Press the
Move the ultrasound probe all the way in the other direction.
Press the
Use the track ball to move to the center of the crosshairs. In the display (F/Fig. 11), a
value of - see Table - must appear that is relative to the probe. If this is not the case, the
adjustment procedure must be repeated for the potentiometer in the ultrasound arm or
the ultrasound arm must be replaced.
Probe
Sector (1/Fig. 2)
Array (2/Fig. 2)
Press the
Siemens AG
Medical Solutions
F
D
The isocenter must be checked with every probe type and after
every replacement of a probe.
key and select the measuring point
Changeover with
key ("Freeze" is deselected).
key and select the measuring point
Changeover with
key ("Freeze" is deselected).
SPL1-130.840.02
Rev. 03
Fig. 11
.
.
Value for D1
15 ± 1 mm
with default values,
see SPL1-130.038.01
13 ± 1 mm
.
Value for D1
153 ± 1 mm
with default values,
see SPL1-130.038.01
151 ± 1 mm
Page 7 of 16
04.05
CS PS 24
6 - 7
F
D
MODULARIS Uro Plus