General Startup

Function check of the EMERGENCY STOP switches

Connect the system to the voltage and boot it.
Click "Emergency" for the patient registration.
Activate the EMERGENCY STOP switch on the C-arm.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
Disengage the EMERGENCY STOP switch.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop"
Siemens AG
Medical Solutions
SPR2-330.815.01.01.02
02.05
(C-arm movement and emergency stop / p.
Page 17 of 46
CS SD 24
17
0
38).