Error Displays Of The Image Memory Interface; Error Displays Of The Shock Wave System Interface; Error Displays Of The Brightness Actual Value Interface - Siemens LITHOSTAR Multiline Troubleshooting Instructions

Hide thumbs Also See for LITHOSTAR Multiline:
Table of Contents

Advertisement

3 - 2

Error displays of the image memory interface:

Error No. Meaning
30
31
32
33
34
35
36
37
38
3A

Error displays of the shock wave system interface:

Error No. Meaning
40
41
42
43
44
45
46
47
48
4A
4E

Error displays of the brightness actual value interface:

Error No. Meaning
50
51
52
53
54
55
56
57
58
5A
LITHOSTAR Multiline
System
No first contact to the HOST (status message)
Timeout error from the HOST
Fatal error in the data reception block from the HOST
Illegal data in the data reception block from the HOST
Illegal value in the data reception flag from the HOST
No first contact to the image memory (status message)
Image memory timeout error
Data reception error from image memory
Serial transmitter not free
Error in the DPR reception area
No first contact to the HOST (status message), System not registered
Timeout error from the HOST
Fatal error in the data reception block from the HOST
Illegal data in the data reception block from the HOST
Illegal value in the data reception flag from the HOST
No first contact to the shock wave system (status message)
Shock wave system timeout error
Data reception error from shock wave system
Serial transmitter not free
Error in the DPR reception area
Respiratory trigger signal timeout
No first contact to the HOST (status message)
Timeout error from the HOST
Fatal error in the data reception block from the HOST
Illegal data in the data reception block from the HOST
Illegal value in the data reception flag from the HOST
No first contact to the brightness actual value (status mes-
sage)
Timeout error from the brightness value actual value signal
device
Data reception error of brightness actual value
Serial transmitter not free
Error in the DPR reception area
Communication controller D6
RXL2-120.072.03
Rev. 04
06.02
Page 2 of 4
CS PS 24
3
3
3
Siemens AG
Medical Solutions