5
Method Parameters
Method Parameters Comprehensive 2D-LC
3 To define Upslope (slope of the rising peak), add the required values to the
4 To define Downslope (slope of the falling peak), add the required values to the
5 To define Threshold (height of the peak that triggers collection), add the
6 To define Upper threshold (height of the peak that ensures that collection is

Gradient Preview Functionality

The gradient preview provides the following functions:
Reference Signal
The user can load a chromatographic signal from an LC detector - a so-called
reference signal. The signal will be shown in the gradient preview. This signal is
automatically shown in the gradient preview of the setup dialog as long as the
reference signal is part of the method. The signal can also be removed from the
method or replaced by another signal.
Loading / removing a reference signal is triggered by toolbar buttons (or the
context menu) of the gradient preview.
2D-LC - User Guide
Threshold only
Detects peaks based on threshold values only.
Threshold/Slope values
Detects peaks based on both - threshold and slope.
Slope only
Detects peaks based on slope values only.
corresponding field.
corresponding field.
required values to the corresponding field.
not switched off even for a saturated signal that might be expected to do so),
add the required values to the corresponding field.
Displays the gradient (%B) of the 1
Displays the gradient (%B) of the 2
dimension, depending on the selected 2D-LC mode (comprehensive OR
heart-cutting)
Allows to graphically edit the gradient shifting
Displays a reference signal by which a user can easily setup the trigger table
or optimize his peak detector settings