5
Method Parameters
Method Parameters Comprehensive 2D-LC

Define Peak Detector Parameter

This section allows parameterizing the peak detector to be used for
peak-triggered 2D-LC operation (comprehensive or heart cutting).
Figure 67
The stop time for a 2D-LC measurement must be set for the 2D pump, which can
be accessed through the advanced settings. It must be at least the 1D run time
and applies to the entire measurement, not to partial 2D-only runs/gradients for
parked peaks.
Multiple Heart-Cutting automatically extends this run time, if required, as
analyzing parked peaks takes usually longer than the 1
If you define a 1D stop time, it will be applied unchanged, for example the
analysis will stop after that time without processing any parked peaks. This is not
recommended and will lead to a warning in the gradient preview.
If no peak detector is configured (see
NOTE
disabled. The currently configured peak detector (name & serial number of the
detector) is shown in the section header.
To facilitate the determination of parameters, it is possible to preview Threshold
NOTE
and Slope in the reference chromatogram.
1 Go to Instrument> Setup 2DLC and tab Advanced.
2 Select Peak detection mode from the drop-down list.
2D-LC - User Guide
Overview on Peak detector parameters
The following options are available:
Off
The peak detector is not used.
st
dimension run only.
"Configuration"
on page 148) this section is
136