14
Agilent 2D-LC Solution with the selectivity of mass selective detection (MSD)
Agilent 2D-LC Solution with high-end MS detection
Agilent 2D-LC Solution with high-end MS
detection
Figure 117 Typical system setup
OpenLab CDS ChemStation and MassHunter have passed the test successfully
NOTE
for co-execution regarding parallel data acquisition. For further information
please check the recommended conditions in the technical note 2D
Chromatogram Creator for MassHunter.
The LC part is an LC system as described in the previous pages of this 2D-LC
system manual. A typical system uses a UV detector in the first dimension and
an optional UV detector in the second dimension. The 2D-LC instrument control
and UV signal acquisition is still done with OpenLab CDS ChemStation (CS)
Edition. Therefoer the 2D-LC software requires configuring at least one UV
detector.
The MS detector, a TOF, Q-TOF, or Triple Quadrupole, is used in the second
dimension. It is controlled by MassHunter (MH) Acquisition software, which
acquires the MS signal.
2D-LC - User Guide
318