10
Error Information
Module-Specific Error Messages

Pressure Cluster Partner Missing

The connection from the valve drive to a defined pressure cluster partner is lost.
Probable cause
1
2

Position Cluster Partner Missing

Probable cause
1
2
2D-LC - User Guide
Communication issues
Configuration mismatch
Communication issues
Configuration mismatch
Suggested actions
Check the CAN cable connections of the mod-
ules.
Check and correct if necessary the valve configu-
ration and presence of defined pressure cluster
partner.
Suggested actions
Check the CAN cable connections of the mod-
ules.
Check and correct if necessary the valve configu-
ration and presence of defined position cluster
partner.