Method Parameters
Method Parameters Comprehensive 2D-LC

Set solvents

The selection of the solvents itself must be done in the standard pump method
NOTE
user interface.
1 Set the percentage of solvent B to any value from 0 – 100 % in steps of
Percentage settings
Solvent information
The corresponding Percent B value in the Standard Pump user interface will be
NOTE
ignored as long as the 2D-LC functionality is enabled (see
page 148).
2D-LC - User Guide
Open the pump method dialog using the button Advanced
and change the selection of the solvents there.
After closing the dialog, the solvent settings should be updated immediately.
0.01 %.
Figure 64
2D-LC solvent settings
Solvent A always delivers the remaining percentage of volume. If the rate of
solvent B is for example set to 20 %, solvent A, following the calculation %A =
100 - %B, automatically is set to 80 %.
The name of the selected solvents and their solvent channels (A1:... or A2:...
and B1:... or B2:...) are shown in the corresponding text fields.
2
D pump settings...
"Configuration"
on
129