External Valve Falls Into Resident Mode - Agilent Technologies 1290 Infinity II 2D-LC Solution User Manual

Hide thumbs Also See for 1290 Infinity II 2D-LC Solution:
Table of Contents

Advertisement

10
Error Information
Module-Specific Error Messages

External Valve falls into resident mode

Error ID: Flashing status indicator
The valve drive was not able to operate correctly
Probable cause
1
2
2D-LC - User Guide
Communication issues
Configuration mismatch
Suggested actions
Check the CAN cable connections of the
modules.
Check if the hosted module is present.
Check if the firmware on the entire stack is
out of the same firmware set.
Check if the limit of 3 hosted modules for
each host module is not exceeded.
Check if the dipswitch settings are correct.
Check if the firmware on the entire stack has
to be the latest version.
257