6.7.2. Analog Output (Step Test)

This test can be used to assure the analog outputs are calibrated and working
properly. The test forces the A1, A2 and A4 analog output channels to produce
signals ranging 0% to 100% of whatever the range in which they are set in 20%
increments. This test is useful for verifying the operation of the datalogging/
recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
Pressing the key under the STEP % will
Brackets will appear Around the Step %:
Pressing the same key again will resume
96
pause the Test.
EXAMPLE [20]
the Test.
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
SETUP X.X
MORE
CFG DAS RNGE PASS CLK
SETUP X.X
DIAG
COMM VARS
HALT
SETUP X.X
ENTER DIAG PASS: 818
8
1
8
DIAG
SIGNAL I / O
NEXT
DIAG
ANALOG OUTPUT
PREV
NEXT
DIAG AOUT
ANALOG OUTPUT
0%
DIAG AOUT
ANALOG OUTPUT
[0%]
Model 400E Ozone Analyzer
Instruction Manual
O3 =XXX.X
SETUP
EXIT
Time to Return
EXIT
Sample Display
ENTR
EXIT
ENTR
EXIT
ENTR
EXIT
Analog Output
EXIT
EXIT
P/N 04316 Rev: B
Exit at Any
to main the
Performs
step test.
0% - 100%
Exit 2X's
to Return
to the
Diag Menu