CALIBRATION PROCEDURES
STEP TWO: Set the expected O
SAMPLE
RANGE = 500.0 PPB
< TST TST >
CAL
M-P CAL
< TST TST > ZERO
M-P CAL
O3 SPAN CONC:
0
0
4
ENTR stores the expected O3
span concentration value.
It is important to independently verify the
172
O3 =XXX.X
RANGE = 500.0 PPB
O3 =XXX.X
CONC
400.0 Conc
0
0
.0
ENTR EXIT
Concentration Value of the SPAN gas.
Span Gas concentration.
3
SETUP
This sequence causes the
M400E to prompt for the
expected O3 span
EXIT
The O3 span concentration value automatically
Make sure that you input the ACTUAL
To change this value to meet the actual
concentration of the SPAN Gas, enter the
number sequence by pressing the key under
each digit until the expected value is set.
EXIT 2x's to Return to the
Main SAMPLE Display
NOTE
Model 400E Ozone Analyzer
Instruction Manual
concentration.
defaults to 400.0 Conc .
concentration value of the SPAN Gas.
PRECISE
O
3
P/N 04316 Rev: B