7.2.3.

DIAG Menu

The DIAG menu (See Figure 7-2 and Figure 7-3) contains functions that are useful for
testing and configuring external equipment that may be connected to the rear panel I/O of
the instrument.
7.2.3.1. Analog Step-Test
When placed in this mode, the instrument will automatically step the analog output
through 5 points from 0 to 100 % and display the current value on the front panel. The
operator can suspend the automatic cycling and manually control the stepping by pressing
either one of the UP / DOWN arrow buttons. To resume automatic cycling, exit the
function by pressing CFG and re-enter by pressing ENT.
The 5 points will correspond to the following analog output values depending on whether
the output is configured for 0-5V or 4-20mA:
Point
0%
25%
50%
75%
100%
Press CFG to exit from the Analog Step-Test function.
7.2.3.2. Signal I/O
The Signal I/O menu allows the operator to manually control the various relay outputs
and control inputs available on the rear panel of the 465M. This function is useful for
testing or debugging external control systems (i.e. dataloggers or PLC's) to which the
instrument may be connected.
See Section 4.5 for additional details on making connections to these I/O signals on the
rear panel.
See Table 7-4 for additional details on entering the Signal I/O menu.
36
Teledyne API Model 465M O

Table 7-3: Analog Step Test Values

0-5V Output
0.00 V ± 0.02
1.25 V ± 0.02
2.50 V ± 0.02
3.75 V ± 0.02
5.00 V ± 0.02
Monitor User Manual
3
4-20 mA Output
4 mA ± 0.02
8 mA ± 0.02
12 mA ± 0.02
16 mA ± 0.02
20 mA ± 0.02
06160J DCN7969