Teledyne 400E Instruction Manual page 293

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
CUSTOMER: ____________________________________________ PHONE: _______________________________________
CONTACT NAME: _______________________________________ FAX NO. _______________________________________
SITE ADDRESS: _________________________________________________________________________________________
MODEL 400E SERIAL NO.: ___________________ FIRMWARE REVISION: ______________________________________
1. ARE THERE ANY FAILURE MESSAGES? _______________________________________________________________
________________________________________________________________________________________________________
________________________________________________________________________________________________________
PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TESTS
PARAMETERS BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT)
2. *IF OPTIONS ARE INSTALLED
Parameter
RANGE
STABIL
O3 MEAS
O3 REF
O3 GEN*
O3 DRIVE*
PRES
SAMPLE FL
SAMPLE TEMP
PHOTO LAMP
O3 GEN TMP*
BOX TEMP
SLOPE
OFFSET
REF_4096_MV
REF_GND
3. WHAT IS THE SAMPLE FLOW AND SAMPLE PRESSURE WITH THE SAMPLE INLET ON REAR OF MACHINE
CAPPED?
SAMPLE FLOW -
4. WHAT ARE THE FAILURE SYMPTONS?_________________________________________________________________
_________________________________________________________________________________________________________
_________________________________________________________________________________________________________
5. IF POSSIBLE, PLEASE INCLUDE A PORTION OF A STRIP CHART PERTAINING TO THE PROBLEM. CIRCLE THE
PERTINENT DATA.
6. THANK YOU FOR PROVIDING THIS INFORMATION. YOUR ASSISTANCE ENABLES TELEDYNE API TO RESPOND
FASTER TO THE PROBLEM THAT YOU ARE ENCOUNTERING.
PHONE: (858) 657- 9800
04404 Rev B
Appendix C
Warranty/Repair Questionnaire
Model 400E
Recorded Value
Values are in the Signal I/O
3
CM
/MIN
SAMPLE PRESSURE -
TELEDYNE API CUSTOMER SERVICE
EMAIL:
api-customerservice@teledyne.com
TOLL FREE: (800) 324-5190
Acceptable Value
PPB/PPM
1 – 10,000 PPB
<= 0.3 PPM WITH ZERO AIR
mV
2500 – 4800 mV
mV
2500 – 4800 mV
mV
80 mV. – 5000 mV.
mV
0 – 5000 mV.
IN-HG-A
~ - 2"AMBIENT ABSOLUTE
3
CM
/MIN
ºC
ºC
50 ºC ± 1 ºC
ºC
48 ºC ± 3 ºC
ºC
PPB
0.0 ± 5.0 PPB
mV
4096mv±2mv and Must be Stable
mV
0± 0.5 and Must be Stable
FAX: (858) 657-9816
TELEDYNE
INSTRUMENTS
Advanced Pollution Instrumentation
A Teledyne Technologies Company
800 ± 10%
10 – 50 ºC
10 – 50 ºC
1.0 ± .15
IN-HG-A
1