Configuring The Internal Zero/Span Option (Izs); Setting The O; Generator Span-Check Output Level - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
6.4. Configuring the Internal Zero/Span Option
(IZS)
In order to use the IZS option to perform calibration checks, it is necessary to
configure certain performance parameters of the O

6.4.1. Setting the O

To set the ozone SPAN concentration for the IZS O
Toggle these
keys to
change
setting
P/N 04316 Rev: B

Generator Span-Check Output Level

3
SAMPLE
RANGE = 500.0 PPB
< TST
CAL
CALZ
SPAN CAL M
RANGE = 500.0 PPM
<TST TST> SPAN
SPAN CAL M
SPAN CONC MENU
SPAN
O3GN
SPAN CAL M
3)
O3_GEN_SET=400.0 PPB
0
4
0
0
• Only Values from 0 to 1500 will be accepted..
• A value of 0 turns the lamp OFF.
The ENTR key will disappear if an invalid setting is
attempted.
Generator.
3
generator, press:
3
O3 =XXX.X
CALS
O3 =XXX.XX
CONC
O3 =XXX.XX
.0
ENTR EXIT
Operating Instructions
SETUP
EXIT
EXIT
EXIT ignores the
new setting.
ENTR accepts the
new setting.
69