Teledyne 400E Instruction Manual page 98

Ozone analyzer
Table of Contents

Advertisement

Operating Instructions
6.7.3.1. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should
only be necessary after major repair such as a replacement of CPU, motherboard or
power supplies. Activate the ANALOG I/O CONFIGURATION MENU, then press:
Instrument
calibrates
automatically
98
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
Continue pressing SET> until ...
AIN CALIBRATED:
DIAG AIO
< SET SET>
CAL
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
DIAG AIO
AIN CALIBRATED:
< SET SET>
CAL
Model 400E Ozone Analyzer
Instruction Manual
ENTR
EXIT
NO
EXIT
YES
EXIT
Exit at any time to
return to the main
DIAG menu
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
P/N 04316 Rev: B