7.3. Manual Calibration CHECKS without Zero/
Span Valve or IZS Options
This is the basic method for manually CHECKING the calibrating the Model 400E O
Analyzer.
STEP ONE: Connect the Sources of Zero Air and Span Gas as shown below.
MODEL 701
Zero Air Generator
Figure 7-2: Pneumatic connections for Manual Calibration without Z/S Valve
P/N 04316 Rev: B
MODEL 700
Gas Dilution Calibrator
(w/ Photometer Option)
or
MODEL 401
Ozone Generator
VENT
or IZS Options
CALIBRATION PROCEDURES
Source of
SAMPLE GAS
Removed during
calibration
Sample
MODEL
Exhaust
400E
Span
Zero Air
Dry Air