Specifications, Agency Approvals, And Warranty; Specifications; Table 2-1: Model 400E Basic Unit Specifications - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
2. SPECIFICATIONS, AGENCY APPROVALS,
AND WARRANTY

2.1. Specifications

Min/Max Range
(Physical Analog Output)
Measurement Units
Zero Noise
Span Noise
Lower Detectable Limit
Zero Drift (24 hours)
Zero Drift (7 days)
Span Drift (24 hours)
Span Drift (7 days)
Linearity
Precision
Lag Time
Rise/Fall Time
Sample Flow Rate
Temperature Range
Humidity Range
Pressure Range
Altitude Range
Temp Coefficient
Voltage Coefficient
Dimensions (H x W x D)
Weight
AC Power
Environmental Conditions
Analog Outputs
Analog Output Ranges
Analog Output Resolution
Status Outputs
Control Inputs
Serial I/O
Certifications
P/N 04316 Rev: B

Table 2-1: Model 400E Basic Unit Specifications

Min: 0-100 PPB
Max: 0-10,000 PPB
ppb, ppm, µg/m
< 0.3 ppb RMS (EPA Definition)
< 0.5% of reading above 100 PPB (EPA Definition)
< 0.6 PPB (EPA Definition)
< 1.0 ppb (at constant temperature and voltage)
< 1.0 ppb (at constant temperature and voltage)
< 1% of reading (at constant temperature and voltage)
< 1% of reading (at constant temperature and voltage)
< 1% of full scale
< 0.5% of reading (EPA Definition)
< 10 sec (EPA Definition)
< 20 sec to 95% (EPA Definition)
800 ± 80 cc/min
5 - 40°C
0-90% RH, Non-Condensing
25 – 31 "Hg-A
0-2000m
< 0.05% per deg C
< 0.05% per Volt AC (RMS) over range of nominal ± 10%
7" x 17" x 23.5"
30.6lbs. (13.8Kg) with IZS Option
100V 50/60Hz (3.25A), 115V 60Hz (3.0A), 220 – 240 V 50/60 Hz (2.5A)
Installation Category (Over voltage Category) II Pollution Degree 2
Four (4) Outputs, Three (3) defined
All Outputs: 100 mV, 1 V, 5 V, 10 V
Two concentration outputs convertible to 4-20 mA isolated current loop
All Ranges with 5% Under/Over Range
1 part in 4096 of selected full-scale voltage
8 Status outputs from opto-isolators
6 Control Inputs, 3 defined, 3 spare
COM1: RS-232; COM2: RS-232 or RS-485
Baud Rate : 300 – 115200
USEPA: Equivalent Method Number EQOA-0992-087
CE Mark
Specifications, Agency Approvals, and
3
3
, mg/m
(user selectable)
Warranty
17

Advertisement

Table of Contents
loading

Table of Contents