Model 400E Ozone Analyzer
Instruction Manual
To make these adjustments the AOUT AutoCal feature must be turned off (see
Section 6.7.3.3) then press:
Press SET> to select the Analog Output
channel to be configured:
DISPLAYED AS =
CONC_OUT_1
CONC_OUT_2
TEST OUTPUT
Then Press EDIT to continue
If AutoCal is O N, go to
Section 6.7.3.5.
These keys
increment/decrement the
ZERO/SPAN D-to-A converter
100, 10 or 1 counts
respectively.
Continue adjustments until the
voltage measured at the
output of the analyzer and/or
the input of the recording
device matches the value in
the upper right hand corner of
the display to the tolerance
listed in Table 6-11.
The analyzer display WILL
NOT CHANGE . Only the
voltage reading of your volt
meter will change.
P/N 04316 Rev: B
FROM ANALOG I/O CONFIGURATION MENU
DIAG
PREV
NEXT
DIAG AIO
< SET
SET>
CHANNEL
=
A1
=
A2
=
A4
DIAG AIO
< SET SET>
DIAG AIO
SET>
DIAG AIO
< SET
SET>
DIAG AIO
< SET
DIAG AIO
< SET
DIAG AIO
output by
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
< SET
ANALOG I / O CONFIGURATION
ENTR
EXIT
AOUTS CALIBRATED: NO
CAL
EXIT
CONC_OUT_1 :5V, NO CAL
EDIT
EXIT
CONC_OUT_1 RANGE: 5V
EDIT
EXIT
CONC_OUT_1 REC OFS: 0 mV
EDIT
EXIT
CONC_OUT_1 AUTO CAL: OFF
SET>
EDIT
CONC_OUT_2 CALIBRATED: NO
CAL
CONC_OUT_1 VOLT–Z : 0 mV
CONC_OUT_1 VOLT–S : 4500 mV
CONC_OUT_1 CALIBRATED: YES
CAL
Operating Instructions
EXIT
EXIT
EXIT ignores the
new setting.
ENTR accepts the
new setting.
EXIT
105