Teledyne 400E Instruction Manual page 101

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
6.7.3.3. Selecting for Auto/Manual Analog Output
The analog outputs configured for voltage mode can be calibrated either
automatically or manually. In its default mode the instrument is configured for
automatic calibration. Manual calibration should be used for the 0.1V range or in
cases where the outputs must be closely matched to the characteristics of recording
device. Outputs configured for automatic calibration can be calibrated as a group or
individually.
To select manual output calibration for a particular channel press the following:
Toggles the Auto
Cal Mode
ON/ OFF for this
Analog Output
channel only.
P/N 04316 Rev: B
Calibration
FROM ANALOG I/O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
SET>
CAL
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
DIAG AIO
CONC_OUT_2 AUTO CAL: ON
< SET
SET> EDIT
DIAG AIO
AOUT AUTO CAL: ON
ON
Exit to Return
to the main
Sample Display
ENTR
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
ENTR EXIT
Operating Instructions
Press SET> to select the Analog Output
channel to be configured:
DISPLAYED AS
=
CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
TEST OUTPUT
=
A4
Then Press EDIT to continue
Pressing ENTR records the new setting
and returns to the previous menu
Pressing EXIT ignores the new setting and
returns to the previous menu
101