APPENDIX A-2: Setup Variables For Serial I/O, Revision C.3
SETUP VARIABLE
O3_GEN_SET2
O3_GEN_DEF
REF_DELAY
REF_FREQ
REF_FSIZE
REF_INTEG
REF_DERIV
DRIVE_STABIL
CACHE_RESOL
O3_SPAN1
O3_SLOPE1
O3_OFFSET1
O3_SPAN2
O3_SLOPE2
O3_OFFSET2
DYN_ZERO
DYN_SPAN
RANGE_MODE
CONC_RANGE1
CONC_RANGE2
RS232_MODE
BAUD_RATE
MODEM_INIT
10
NUMERIC
DEFAULT
UNITS
VALUE
PPB
400
PPB
400
Seconds
60
Seconds
12
Samples
4
0.1
0
mV
10
PPB
2
Conc
400
1
PPB
0
Conc
400
1
PPB
0
OFF
OFF
SNGL
0
Conc
500
Conc
500
BitFlag
0
0
19200
"AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0
E0 Q1
0
&W0"
Model 400E Instruction Manual
VALUE RANGE
0–1500
O
generator high set point for range #2.
3
0–1500
O
generator default set point.
3
1–300
Delay before beginning O
reference feedback control.
1–60
O
generator reference adjustment
3
frequency.
1–10
O
generator reference filter size.
3
0–10
O
generator reference PID integral
3
coefficient.
0–10
O
generator reference PID derivative
3
coefficient.
0.1–100
O
generator drive stability limit for
3
concentration cache updates.
0.1–20
O
generator cache un-normalized
3
concentration resolution.
50–10000
Target O
calibration for range #1.
0.850–1.150
O
slope for range #1.
3
-100–100
O
offset for range #1.
3
50–10000
Target O
calibration for range #2.
0.850–1.150
O
slope for range #2.
3
-100–100
O
offset for range #2.
3
OFF, ON
ON enables dynamic zero calibration for
contact closures and Hessen protocol.
OFF disables it.
OFF, ON
ON enables dynamic span calibration for
contact closures and Hessen protocol.
OFF disables it.
SNGL, DUAL, AUTO
Range control mode.
0.1–20000
D/A concentration range #1.
0.1–20000
D/A concentration range #2.
0–65535
RS-232 COM1 mode flags. Add values to
combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
16 = enable Hessen protocol
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line errors
256 = disable XON / XOFF support
512 = disable hardware FIFOs
1024 = enable RS-485 mode
2048 = even parity, 7 data bits, 1 stop
bit
4096 = enable command prompt
300, 1200, 2400,
RS-232 COM1 baud rate.
4800, 9600,
19200, 38400,
57600, 115200
Any character in
RS-232 COM1 modem initialization
the allowed
string. Sent verbatim plus carriage return
character set. Up
to modem on power up or manually.
to 100 characters
long.
DESCRIPTION
generator
3
concentration during span
3
concentration during span
3
5
04402 Rev C.3