EPA Protocol Calibration
Calibration
Activities
Zero-air
Calibrator
Multipoint
202
Table 8-4: Activity Matrix for Calibration Procedures
Acceptance Limits
Zero-air, free of
contaminants
(Sec. 2.0.7 Ref. 11.).
Meet all requirement for
UV photometer as
specified in Sec. 2.7.2
QA Manual, TAD
1
Fed. Reg.
or approve
Transfer Standard
Sec. 2.7.1, Q.A. Manual
3
and TAD
.
According to Calibration
procedure (Sec. 2.7.2
Q.A.. Manual Ref 11) and
Federal Register; data
recorded.
Frequency and
Method o f
Measurement
Compare the new
Zero-air against
Source known to be
free of contaminants.
Re-certify transfer
Standard against
Primary UV
2
and the
Photometer at least
Twice each quarter.
Calibrate at least
Once, quarterly;
Anytime an audit
Indicates
discrepancy; After
maintenance that May
affect the Calibration
(Subsec 2.1) Federal
1
Register
.
Model 400E Ozone Analyzer
Instruction Manual
Action if
Requirements are
not Met
Take corrective action
with generation system
as appropriate.
Return to supplier, or
take corrective action
with system as
appropriate.
Repeat the calibration.
P/N 04316 Rev: B