Teledyne 400E Instruction Manual page 167

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
STEP TWO: Set the expected O
SAMPLE
< TST TST >
CAL
M-P CAL
< TST TST > ZERO
M-P CAL
4
0
0
ENTR stores the expected
CO span concentration value.
For this Initial Calibration it is important to independently verify
P/N 04316 Rev: B
RANGE = 500.0 PPB
RANGE = 500.0 PPB
CONC
O3 SPAN CONC:
400.0 Conc
0
0
.0
ENTR EXIT
the PRECISE O
Concentration Value of the SPAN gas.
3
Span Gas concentration:
3
O3 =XXX.X
SETUP
This sequence causes the
M300E to prompt for the
O3 =XXX.X
EXIT
The O3 span concentration value automatically
number sequence by pressing the key under
EXIT 2x's to Return to the
Main SAMPLE Display
NOTE
CALIBRATION PROCEDURES
expected CO span
concentration.
defaults to 400.0 Conc .
Make sure that you input the ACTUAL
concentration value of the SPAN Gas.
To change this value to meet the actual
concentration of the SPAN Gas, enter the
each digit until the expected value is set.
167