Teledyne 400E Instruction Manual page 279

Ozone analyzer
Table of Contents

Advertisement

Model 400E Instruction Manual
SETUP VARIABLE
RS232_MODE2
BAUD_RATE2
MODEM_INIT2
RS232_PASS
MACHINE_ID
COMMAND_PROMPT
TEST_CHAN_ID
REMOTE_CAL_MODE
PASS_ENABLE
PHOTO_LAMP_POWER
LAMP_PWR_ENABLE
LAMP_PWR_PERIOD
LAMP_OFF_DELAY
DET_VALID_DELAY
REF_SDEV_LIMIT
PHOTO_CYCLE
PHOTO_PROP
PHOTO_INTEG
PHOTO_DERIV
04402 Rev C.3
APPENDIX A-2: Setup Variables For Serial I/O, Revision C.3
NUMERIC
DEFAULT
UNITS
VALUE
0
19200
0
"AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0
E0 Q1
0
&W0"
Password
940331
ID
400
"Cmd> "
0
0
NONE
0
LOW
OFF
mV
4500
OFF
Hours
24
Seconds
0.1
Seconds
20
mV
3
Seconds
5
0.5
0.1
0
VALUE RANGE
0–65535
RS-232 COM2 mode flags.
(Same settings as RS232_MODE.)
300, 1200, 2400,
RS-232 COM2 baud rate.
4800, 9600,
19200, 38400,
57600, 115200
Any character in
RS-232 COM2 modem initialization
the allowed
string. Sent verbatim plus carriage return
character set. Up
to modem on power up or manually.
to 100 characters
long.
0–999999
RS-232 log on password.
0–9999 (Hessen:
Unique ID number for instrument.
0–999)
Any character in
RS-232 interface command prompt.
the allowed
Displayed only if enabled with
character set. Up
RS232_MODE variable.
to 100 characters
long.
NONE,
Diagnostic analog output ID.
PHOTO MEAS,
PHOTO REF,
O3 GEN REF,
SAMPLE
PRESSURE,
SAMPLE FLOW,
SAMPLE TEMP,
PHOTO LAMP
TEMP,
O3 SCRUB TEMP,
O3 LAMP TEMP,
CHASSIS TEMP
LOW,
Range to calibrate during contact closure
or Hessen calibration.
HIGH
OFF, ON
ON enables passwords.
OFF disables them.
0–5000
Photometer lamp power setting.
OFF, ON
ON enables photometer lamp power
cycling.
OFF disables it.
0.01–1000
Photometer lamp power cycling period.
0.02–5
Length of time photometer lamp is
turned off.
1–300
Delay until valid concentration is
computed.
0.1–100
Photometer reference standard deviation
must be below this limit to switch out of
startup mode.
0.5–30
Photometer lamp temperature control
cycle period.
0–10
Photometer lamp temperature PID
proportional coefficient.
0–10
Photometer lamp temperature PID
integral coefficient.
0–10
Photometer lamp temperature PID
derivative coefficient.
DESCRIPTION
11

Advertisement

Table of Contents
loading