Teledyne HASTINGS 200 Series Instruction Manual

Teledyne HASTINGS 200 Series Instruction Manual

Fast response mass flowmeters
Hide thumbs Also See for HASTINGS 200 Series:

Advertisement

Quick Links

TELEDYNE HASTINGS
INSTRUCTION MANUAL
HASTINGS 200 SERIES
HASTINGS 200 SERIES
HASTINGS 200 SERIES
HASTINGS 200 SERIES
HASTINGS 200 SERIES
F F F F F AST RESPONSE
AST RESPONSE
AST RESPONSE
AST RESPONSE
AST RESPONSE
MASS FLOWMETERS
MASS FLOWMETERS
MASS FLOWMETERS
MASS FLOWMETERS
MASS FLOWMETERS
INSTRUMENTS
126B - PAGE 1
INSTRUCTION

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne HASTINGS 200 Series

  • Page 1 TELEDYNE HASTINGS INSTRUCTION INSTRUMENTS INSTRUCTION MANUAL HASTINGS 200 SERIES HASTINGS 200 SERIES HASTINGS 200 SERIES HASTINGS 200 SERIES HASTINGS 200 SERIES F F F F F AST RESPONSE AST RESPONSE AST RESPONSE AST RESPONSE AST RESPONSE MASS FLOWMETERS MASS FLOWMETERS...
  • Page 2 Revision B (Document Number 126-082000) ............August 2000 Revision C (Document Number 126-082005) ............August 2005 Visit www.teledyne-hi.com for WEEE disposal guidance. Hastings Instruments reserves the right to change or modify the design of its equipment without any obligation to provide notification of change or intent to change.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS SECTION 1.0 GENERAL INFORMATION....................5 1.1 Features ....................... 5 1.2 Specifications ...................... 5 1.3 Accessories ......................6 1.3.1 Power Supplies ....................6 1.3.2 Alarms ......................6 1.3.3 Totalizer ......................6 1.3.4 4-20 mA Current Converter ................6 SECTION 2.0 INSTALLATION AND OPERATION .................
  • Page 4 4.6 Printed Circuit Board Replacement ..............15 4.7 Sensor Replacement .................... 15 4.8 Troubleshooting ....................15 SECTION 5.0 REPLACEMENT PARTS ....................17 SECTION 6.0 WARRANTY AND REPAIR ....................19 SECTION 7.0 DIAGRAMS AND DRAWINGS ..................21 AssemblyDiagram,PC810FastReponse................22 Schematic, PC-810 Fast Response .................23 HFM-200 Outline Drawng....................
  • Page 5: General Information

    1.0 GENERAL INFORMATION Hastings Mass Flowmeters are designed to accurately measure mass flow without corrections or compensations for gas pressure and temperature. Due to a linear electrical output signal, the flowmeters are ideal for use with totalizers and recorders. Hastings Mass Flowmeters do not require any periodic maintenance under normal operating conditions with clean gases.
  • Page 6: Accessories

    1.3 Accessor Accessor Accessor Accessories Accessor 1.3.1 P 1.3.1 P 1.3.1 Po o o o o w w w w w er Supplies er Supplies er Supplies 1.3.1 P 1.3.1 P er Supplies er Supplies Hastings’ Power Supplies are designed as combination power supplies and digital readout monitors. They can simultaneously power a combination of different Hastings Flow Instruments.
  • Page 7: Installation And Operation

    2.0 INSTALLATION AND OPERATION This section is designed to assist in getting a new flowmeter into operation as quickly and easily as possible. Please read the following thoroughly before attempting to install the instrument. 2.1 Receiving Inspection: 2.1 Receiving Inspection: 2.1 Receiving Inspection: 2.1 Receiving Inspection: 2.1 Receiving Inspection:...
  • Page 8: Electrical Connections

    The standard inlet and outlet fittings for the HFM-229B Flowmeter are NPT thread in the size of the pipe listed on the drawing on page 22, for sizes up to 3" laminar. On the 4", 6", and 8" laminars, flange fittings are standard.
  • Page 9: Theory Of Operation

    3.0 THEORY OF OPERATION This section contains an overall functional description of Model HFM Flowmeters. Detailed schematics and parts lists can be found at the end of the manual in Section 6.0. In this section and other sections throughout this manual, when a power supply is mentioned, it is assumed that the customer has a Hastings Power Supply.
  • Page 10: Electronics

    For a constant power input, the differential thermocouple output is a function of the mass flow rate and the heat capacity of the gas. Since the heat capacity of many gases is relatively constant over wide ranges of temperature and pressure, the flowmeter may be calibrated directly in mass units for those gases.
  • Page 11 The HFM-230 uses a shunt which is a corrugated and fused shunt similar to the shunt used in the lower flow range instruments (see Figure 3.7). This highest range shunt is adjustable from 0- 50 slpm to 0-500 slpm. The HFM-229B Series uses an external laminar flow element instead of an internal shunt, because the necessary size for the element exceeds the body size of the flowmeter.
  • Page 12 126B - PAGE 12...
  • Page 13: Section 4.0 Maintenance

    4.0 MAINTENANCE 4.1 Introduction: 4.1 Introduction: 4.1 Introduction: 4.1 Introduction: 4.1 Introduction: This section contains service and calibration information. Some portions of the instrument are delicate. Use extreme care when servicing the flowmeter. The potentiometer positions and the electrical compo- nents referred to in the troubleshooting section can be found in Section 6.0 on the electrical component layout drawing.
  • Page 14: End Cap Removal

    End Cap Removal End Cap Removal End Cap Removal End Cap Removal End Cap Removal The end cap on the inlet side must be removed to gain access to the filter or shunt assembly. First shut off the supply of gas to the instrument. Disconnect the Swagelok fitting on the inlet and outlet sides of the instrument.
  • Page 15: Printed Circuit Board Replacement

    4.6 Printed Circuit Board 4.6 Printed Circuit Board 4.6 Printed Circuit Board 4.6 Printed Circuit Board 4.6 Printed Circuit Board Replacement Replacement Replacement Replacement Replacement In the unlikely event that the PC board fails, it is easily removed from the instru- ment and replaced with a spare to minimize instrument downtime.
  • Page 16 SYMPTOM: SYMPTOM: SYMPTOM: SYMPTOM: SYMPTOM: Sensor has proper resistance readings, but little or no output with flow. CAUSE: CAUSE: CAUSE: Plugged sensor. CAUSE: CAUSE: A A A A A CTION: CTION: CTION: CTION: CTION: Shut off gas supply and power supply. Remove cover and PC board from unit. Remove sensor assembly and examine.
  • Page 17: Section 5.0 Replacement Parts

    5.0 REPLACEMENT PARTS The following is a list of the available replacement parts and their factory stock numbers. The HFM- 229, HFM-229B, HFM-200, HFM-200B, and the HFC-202 shunts and filter discs are interchange- able. The HFM-230, HFM-201, and the HFC-203 shunts and filter discs are also interchangeable. The same sensor module is used on all of the above models.
  • Page 18 126B - PAGE 18...
  • Page 19: Section 6.0 Warranty And Repair

    If you then choose not to have the product repaired, a minimum will be charged to cover the processing and inspection. Please consult the factory for your RMA number before returning any product for repair. TELEDYNE HASTINGS 804 NEWCOMBE AVENUE HAMPTON, VIRGINIA 23669 U.S.A.
  • Page 20 126B - PAGE 20...
  • Page 21: Section 7.0 Diagrams And Drawings

    7.0 Diagrams and Drawings 126B - PAGE 21...
  • Page 22: Assemblydiagram,Pc810Fastreponse

    PC-810 Assembly 126B - PAGE 22 (Information from Hastings drawing 30837 rev. F)
  • Page 23: Schematic, Pc-810 Fast Response

    PC-810 Schematic 126B - PAGE 23 (Information from Hastings drawing 30837 rev. F)
  • Page 24: Hfm-200 Outline Drawng

    126B - PAGE 24...
  • Page 25: Hfm-201 Outline Drawing

    126B - PAGE 25...
  • Page 26: Hfm-200B Outline Drawing

    126B - PAGE 26...
  • Page 27: Laminar Flow Element Outline Drawing

    126B - PAGE 27...
  • Page 28 126B - PAGE 28...