4.13.3. ANALOG OUTPUT STEP TEST

SAMPLE
< TST TST > CAL
SETUP X.X
CFG DAS RNGE PASS CLK
SETUP X.X
COMM VARS
SETUP X.X
DIAG
NEXT
DIAG
PREV
DIAG AOUT
DIAG AOUT
Pressing the key under "0%" while performing the test will
128
This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to
100% of the full scale range in 20% increments. This test is useful to verify the
operation of the data logging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
A1:NXCNC1=100PPM
PRIMARY SETUP MENU
SECONDARY SETUP MENU
DIAG
ENTER DIAG PASS: 818
8
1
8
SIGNAL I / O
ENTR
ANALOG OUTPUT
NEXT
ANALOG OUTPUT
0%
ANALOG OUTPUT
[0%]
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again
will resume the test.
Teledyne API - Model T200H/T200M Operation Manual
NOX=XXX.X
SETUP
MORE
EXIT
EXIT
ENTR
EXIT
EXIT
ENTR
EXIT
EXIT
EXIT
Performs
analog output
step test.
0% - 100%
Exit-Exit
returns to the
DIAG menu
07270D DCN7141