7.2.3. DIAG Menu

The DIAG menu (see Figure 7-2) contains functions that are useful for testing and
configuring external equipment that may be connected to the rear panel I/O of the
instrument.
7.2.3.1. Analog Step-Test
When placed in this mode, the instrument will automatically step the analog output
(See Section 4.4.2 for details on the analog output connection) through 5 points
from 0 to 100 % and display the current value on the front panel. The operator can
suspend the automatic cycling and manually control the stepping by pressing either
one of the UP / DOWN arrow buttons. To resume automatic cycling, exit the
function by pressing CFG and re-enter by pressing ENT.
The 5 points will correspond to the following analog output values depending on
whether the output is configured for 0-5V or 4-20mA:
Press CFG to exit from the Analog Step-Test function.
05509G DCN6806

Table 7-3: Analog Step Test Values

Point
0-5V Output
0%
0.00 V
25%
1.25 V
50%
2.50 V
75%
3.75 V
100%
5.00 V
Teledyne API Model 465L O
Teledyne API Model 465L O 3 Monitor Operation Manual
4-20 mA Output
4 mA
8 mA
12 mA
16 mA
20 mA