STEP TWO: Perform the Zero/Span Check. Zero and Span checks using the
Zero/Span Valve option are similar to that described in Section 7.3, except that:
On instruments with IZS options, Zero Air and Span Gas are supplied to the
instrument by an internal O
On instruments with Z/S Valve Options, Zero Air and Span Gas is supplied to
the analyzer through the Zero Gas and Span Gas inlets.
The Zero And Cal operations are initiated directly and independently with
dedicated keys (CALZ & CALS).
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the ZERO button
is displayed
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the SPAN button
is displayed
P/N 04316 Rev: B
generator.
3
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
CALZ
CALS
ZERO CAL M
RANGE = 500.0 PPB
ZERO
< TST TST >
CONC
SAMPLE
RANGE = 500.0 PPB
CALS
< TST TST > CAL
CALZ
ZERO CAL M
RANGE = 500.0 PPB
< TST TST >
ZERO
SPAN
CALIBRATION PROCEDURES
O3 =XXX.X
SETUP
O3 =XXX.X
EXIT
O3 =XXX.X
SETUP
O3 =XXX.X
CONC
EXIT
Record the O
3
reading as
displayed on the
instrument's front
panel
Record the O
3
reading as
displayed on the
instrument's front
panel
Press EXIT to Return to the
Main SAMPLE Display
177