Teledyne 400E Instruction Manual page 76

Ozone analyzer
Table of Contents

Advertisement

Operating Instructions
To set the ranges press following keystroke sequence:
The LOW and HIGH
Ranges have separate
slopes and offsets for
computing the carbon
monoxide concentration.
The two ranges must be
independently calibrated.
.
76
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
SETUP X.X
CFG DAS
RNGE
PASS CLK MORE
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
SETUP X.X
RANGE MODE:
SNGL
DUAL
AUTO
SETUP X.X
RANGE MODE:
SNGL DUAL AUTO
SETUP X.X
RANGE CONTROL MENU
SET
MODE
UNIT
500.0 Conc
SETUP X.X
LOW RANGE:
0
0
1
0
0
SETUP X.X
HIGH RANGE:
500.0 Conc
0
0
5
0
0
SETUP X.X
RANGE CONTROL MENU
MODE SET UNIT
Model 400E Ozone Analyzer
O3 =XXX.X
SETUP
EXIT
EXIT
SNGL
ENTR EXIT
DUAL
ENTR
EXIT
EXIT
.0
ENTR
EXIT
Numeral Keys
to set the upper
limit of each
.0
ENTR
EXIT
EXIT
Instruction Manual
Toggle the
range.
EXIT Returns
to the Main
SAMPLE Display
P/N 04316 Rev: B