Teledyne 400E Instruction Manual page 297

Ozone analyzer
Table of Contents

Advertisement

1
D
SCRUBBER_THCOUPLE_IN
C
B
METAL SCRUBBER HTR
A
1
2
3
P27
CN0000219
MTH+
J18-1
CN0000052
1
+
MTH-
J18-2
THCOUPL1 (K TYPE)
2
-
P10
J19-1
+
J19-2
THCOUPL2
-
(K TYPE, BTM)
THERMIST OUT
THERMIST OUT
JP2-1
TS0
JP2-2
NEU0
JP2-3
TS1
JP2-4
NEU1
JP2-5
TS2
JP2-6
J20-1
NEU2
RL3
JP2-7
J20-2
LD0
NEUT
JP2-8
J20-3
CMN0
RL4
JP2-9
J20-4
LD1
NEUT
JP2-10
J20-5
CMN1
JP2-11
J20-6
LD2
JP2-12
CMN2
J2-1
1
CMN0
P36
J2-2
2
LD0
J2-3
3
TS0
J2-4
4
RLY0
J2-5
5
NEU0
J2-6
6
CMN1
J2-7
7
LD1
J2-8
8
TS1
J2-9
9
RLY1
J2-10
10
NEU1
J2-11
11
CMN2
J2-12
12
LD2
J2-13
13
TS2
J2-14
14
RLY2
J2-15
15
NEU2
J2-16
16
CN0000115
AC-IN
ONLY USED FOR METAL WOOL SCRUBBER OPTION
CN0000052
P12
RELAY_AC_IN
ZERO/SPAN_VALVE
MEAS/REF_VALVE
SAMP/CAL_VALVE
2
3
4
P22
CN0000499
FLOW_PWR
DISPLAY
P18
CN0000499
Relay Power
Connector Pinouts:
pin - Signal
CN0000499
1 - DGND
P20
2 - +5V
3 - AGND
4 - +15V
5 - AGND
6 - -15V
7 - +12RET
8 - +12V
9 - EGND
10 - CGND
MOM_PWR
CN0000499
P23
UV_DET_PWR
CN0000499
P24
VALVE A
P40
CN0000054
P21
CN0000499
4
5
PHOT_LAMP_PWR
J13-2
J13-1
PHOTO_LAMP_HTR
J14-2
J14-1
O3_GEN_HTR
P37
J3-5
JGND
5
J3-4
4
J3-3
INT'
3
J3-2
RSDA
SDA
2
J3-1
RSCL
SCL
1
J6-10
J6-9
J6-8
J6-7
J6-6
J6-5
J6-4
J6-3
J6-2
J6-1
CHASSIS_FAN
J5-10
10
J5-9
9
J5-8
8
J5-7
7
J5-6
6
J5-5
5
J5-4
4
J5-3
3
J5-2
2
J5-1
1
SINGLE_SUPPLY
TRIPLE_SUPPLY
Title
INTERCONNECT DIAGRAM, M400E
Size
Number
04396
Tabloid
Date:
9-Jul-2002
File:
N:\Klwork\400E\M400E-BLK.DDB
5
6
D
C
B
IZS_LAMP_PWR
A
Revision
A
Sheet of
2
2
Drawn By:
6

Advertisement

Table of Contents
loading

Table of Contents