6.7.3.2. Voltage Output Range Selection and Offset
The final step in configuring the analyzer's three analog output channels is to set
the electronic signal type and range of each channel. This consists of selecting a
type, voltage or current (if an optional current output driver has been installed),
and a signal level that matches the input requirements of the recording device
attached to the channel. A bipolar offset can be also added to the signal if required.
Usually this has step has been completed at the factory, but should you need to
change the settings to match a different recording device or as a field upgrade from
voltage to currently loop output, press:
Toggles the Auto
Cal Mode
ON/ OFF for this
Analog Output
channel only.
100
Adjustment
FROM ANALOG I/O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED: NO
< SET
SET>
CAL
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
CONC_OUT_2 RANGE: 5V
DIAG AIO
SET>
EDIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
CONC_OUT_2 AUTO CAL: ON
DIAG AIO
< SET
SET> EDIT
DIAG AIO
AOUT AUTO CAL: ON
ON
Exit to Return
to the main
Sample Display
ENTR
EXIT
Press SET> to select the Analog Output
EXIT
EXIT
EXIT
EXIT
EXIT
ENTR EXIT
Pressing EXIT ignores the new setting and
Model 400E Ozone Analyzer
Instruction Manual
channel to be configured:
DISPLAYED AS
=
CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
TEST OUTPUT
=
A4
Then Press EDIT to continue
Pressing ENTR records the new setting
and returns to the previous menu
returns to the previous menu
P/N 04316 Rev: B