APPENDIX A-5: M400E iDAS Functions, Revision C.3
NAME
PHMEAS
PHREF
SLOPE1
SLOPE2
OFSET1
OFSET2
ZSCNC1
ZSCNC2
CONC1
CONC2
STABIL
O3REF
O3DRIV
O3TEMP
O3STMP
O3SDTY
PHTEMP
PHLDTY
SMPTMP
SMPFLW
SMPPRS
BOXTMP
TEST7
TEST8
TEMP6
TEMP7
REFGND
RF4096
20
Table A-6:
M400E iDAS Functions, Revision C.3
Photometer detector measure reading
Photometer detector reference reading
Slope for range #1
Slope for range #2
Offset for range #1
Offset for range #2
Concentration for range #1 during zero/span calibration, just
before computing new slope and offset
Concentration for range #2 during zero/span calibration, just
before computing new slope and offset
Concentration for range #1
Concentration for range #2
Concentration stability
Ozone generator reference detector reading
Ozone generator lamp drive
Ozone generator lamp temperature
Ozone scrubber temperature
Ozone scrubber temperature duty cycle
Photometer lamp temperature
Photometer lamp temperature duty cycle
Sample temperature
Sample flow rate
Sample pressure
Internal box temperature
Diagnostic test input (TEST_INPUT_7)
Diagnostic test input (TEST_INPUT_8)
Diagnostic temperature input (TEMP_INPUT_6)
Diagnostic temperature input (TEMP_INPUT_7)
Ground reference
Precision 4.096 mV reference
DESCRIPTION
Model 400E Instruction Manual
Degrees C
Degrees C
(1.0 = 100%)
Degrees C
(1.0 = 100%)
Degrees C
Inches Hg
Degrees C
Degrees C
Degrees C
04402 Rev C.3
UNITS
mV
mV
PPB
PPB
PPB
PPB
PPB
PPB
PPB
mV
mV
Fraction
Fraction
cc/m
mV
mV
mV
mV