Teledyne 400E Instruction Manual page 90

Ozone analyzer
Table of Contents

Advertisement

Operating Instructions
6.6.2.3. Analog Output Offset Adjustment
Some analog signal recorders require that the zero signal is significantly different
from the baseline of the recorder in order to record slightly negative readings from
noise around the zero point. This can be achieved in the M400E by defining a zero
offset, a small voltage (e.g., 10% of span), which can be added to the signal of
individual output channels by activating the ANALOG I/O
CONFIGURATION MENU, then pressing:
Set the recorder
offset (in mV) of
the selected
channel
6.6.2.4. Current Loop Output Adjustment
A current loop option is available and can be installed as a retrofit for each of the
analog outputs of the analyzer. This option converts the DC voltage analog output
to a current signal with 0-20 mA output current. The outputs can be scaled to any
set of limits within that 0-20 mA range. However, most current loop applications
call for either 2-20 mA or 4-20 mA range. All current loop outputs have a +5%
over-range. Ranges with the lower limit set to more than 1 mA (e.g., 2-20 or 4-20
mA) also have a -5% under-range.
To switch an analog output from voltage to current loop after installing the current
output printed circuit assembly, follow the instructions in Section 6.9.4.1 and select
CURR from the list of options on the "Output Range" menu.
90
FROM ANALOG I/O CONFIGURATION MENU
ANALOG I / O CONFIGURATION
DIAG
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED: NO
SET>
< SET
CAL
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
DIAG AIO
RECORD OFFSET: 0 MV
+
0
0
0
0
Model 400E Ozone Analyzer
ENTR
EXIT
EXIT
Press SET> to select the
analog output channel to
be configured. Then press
EDIT to continue
EXIT
EXIT
Pressing ENTR accepts the
EXIT
new setting and returns to the
Pressing EXIT ignores the new
ENTR EXIT
Instruction Manual
previous menu.
setting and returns to the
previous menu.
P/N 04316 Rev: B