Model 400E Ozone Analyzer
Instruction Manual
To adjust the a zero and span signal levels of the Current Outputs, press:
These keys
increment/decrement the
ZERO/SPAN D-to-A converter
output by
100, 10 or 1 counts
respectively.
The resulting change in output
voltage is dispalyed on the
TOP Line.
Continue adjustments until the
correct current level is
measured on a current meter
placed in series with the
output between the Model
300E and the recording
device.
P/N 04316 Rev: B
FROM ANALOG I/O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET>
EDIT
DIAG AIO
AIN CALIBRATED: NO
SET>
EDIT
DIAG AIO
AOUT CALIBRATED: NO
< SET
SET>
CAL
DIAG AIO
CONC_OUT_CURR, NO CAL
EDIT
< SET SET>
DIAG AIO
CONC_OUT_2 RANGE: CURR
<SET
SET>
EDIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
DIAG AIO
CONC_OUT_2 ZERO: 0 mV
U100
UP10
UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
CONC_OUT_2 ZERO: 27 mV
U100 UP10 UP DOWN DN10 D100
DIAG AIO
CONC_OUT_2 SPAN: 10000 mV
U100 UP10 UP DOWN DN10
DIAG AIO
CONC_OUT_2 ZERO: 9731 mV
U100 UP10 UP DOWN DN10 D100
DIAG AIO
CONC_OUT_2 CALIBRATED: YES
< SET
CAL
Operating Instructions
Exit to Return
to the main
EXIT
Sample Display
EXIT
EXIT
Press SET> to select the Analog Output
EXIT
channel to be configured:
DISPLAYED AS
CONC_OUT_1
CONC_OUT_2
TEST OUTPUT
Then Press EDIT to continue
EXIT
EXIT
ENTR returns
to the previous
EXIT
EXAMPLE
ENTR EXIT
EXIT ignores the
D100
ENTR EXIT
ENTR accepts the
EXAMPLE
ENTR EXIT
EXIT
=
CHANNEL
=
A1
=
A2
=
A4
menu
new setting.
new setting.
107