Analog Output Step Test - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Model 100E Instruction Manual
To enter the signal I/O test mode, press:
SAMPLE
EXIT returns
to the main
SAMPLE display

6.7.3. Analog Output Step Test

This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to
100% of the full scale range in 20% increments. This test is useful to verify the operation
of the data logging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
SAMPLE
RANGE = 500.000 PPB
< TST TST > CAL
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
SETUP X.X
ENTER DIAG PASS: 818
8
1
8
045150102 Rev XB1
RANGE = 500.000 PPB
SO2 =X.XXX
< TST TST > CAL
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
SETU`P X.X
ENTER DIAG PASS: 818
8
1
8
ENTR
SO2 =X.XXX
SETUP
EXIT
MORE
EXIT
EXIT
ENTR
DIAG
SIGNAL I / O
JUMP
SETUP
PREV NEXT
DIAG I / O
Test Signals Displayed Here
EXIT
PREV NEXT
JUMP
EXAMPLE
DIAG I / O
JUMP TO: 12
EXIT
1
2
DIAG I / O
ST_CONC_VALID = ON
PREV NEXT JUMP
EXIT
Pressing the PRNT key will send a formatted printout to the serial port and can be
captured with a computer or other output device.
DIAG
SIGNAL I / O
NEXT
DIAG
ANALOG OUTPUT
PREV
NEXT
ANALOG OUTPUT
DIAG AOUT
0%
DIAG AOUT
ANALOG OUTPUT
[0%]
Pressing the key under "0%" while performing the test will
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again
will resume the test.
Operating Instructions
Use the NEXT & PREV
keys to move between
EXIT
signal types.
ENTR
Use the JUMP key to
go directly to a
specific signal
See Appendix A-4 for
PRNT EXIT
a complete list of
available SIGNALS
Enter 12 to Jump
to Signal 12:
(ST_CONC_VALID)
EXIT
ENTR
Exit to return
to the
PRNT EXIT
DIAG menu
ON
ENTR
EXIT
EXIT
ENTR
Performs
analog output
step test.
EXIT
0% - 100%
returns to the
DIAG menu
EXIT
Exit-Exit
75

Advertisement

Table of Contents

Troubleshooting

loading