Teledyne 400E Instruction Manual page 86

Ozone analyzer
Table of Contents

Advertisement

Operating Instructions
To automatically calibrate an single analog channel, activate the ANALOG I/O
CONFIGURATION MENU, then press:
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AOUTS CALIBRATED: NO
<
SET>
CAL
DIAG AIO
CONC_OUT_2:5V, CAL
EDIT
< SET SET>
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
CONC_OUT_2 AUTO CAL: ON
DIAG AIO
< SET
SET>
EDIT
86
ENTR
EXIT
EXIT
EXIT
EXIT
EXIT
EXIT
Model 400E Ozone Analyzer
EXIT to Return
to the main
Sample Display
Press SET> to select the
Analog Output channel to
be configured. Then Press
EDIT to continue
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
<SET
CAL
DIAG AIO
AUTO CALIBRATING CONC_OUT_2
DIAG AIO
CONC_OUT_2 CALIBRATED:
<SET
CAL
Instruction Manual
EXIT
YES
EXIT
P/N 04316 Rev: B