APPENDIX A-3: Warnings and Test Functions, Revision C.3
Table A-3:
M400E Test Functions, Revision c.3
Message Text
RANGE=500.0 PPB
RANGE1=500.0 PPB
RANGE2=500.0 PPB
STABIL=0.0 PPB
O3 MEAS=2993.8 MV
O3 REF=3000.0 MV
O3 GEN=4250.0 MV
O3 DRIVE=0.0 MV
PHOTO POWER=4500.0 MV
PRES=29.9 IN-HG-A
SAMP FL=700 CC/M
SAMPLE TEMP=31.2 °C
PHOTO LAMP=52.3 °C
O3 SCRUB=110.2 °C
O3 GEN TMP=48.5 °C
BOX TEMP=31.2 °C
SLOPE=1.000
OFFSET=0.0 PPB
O3=191.6 PPB
TEST=2753.9 MV
TIME=14:48:01
D/A range in single or auto-range modes.
D/A #1 range in dual range mode.
D/A #2 range in dual range mode.
Concentration stability (standard deviation based
3
on setting of STABIL_FREQ and
STABIL_SAMPLES).
Photometer detector measure reading.
Photometer detector reference reading.
O3 generator reference detector reading.
O3 generator lamp drive output.
Photometer lamp drive output.
Sample pressure.
Sample flow rate.
Sample temperature.
Photometer lamp temperature.
O3 scrubber temperature.
O3 generator lamp temperature.
Internal chassis temperature.
Slope for current range, computed during
zero/span calibration.
Offset for current range, computed during
zero/span calibration.
O3 concentration for current range.
Value output to TEST_OUTPUT analog output,
selected with TEST_CHAN_ID variable.
Current instrument time of day clock.
DESCRIPTION
15