Epa Equivalency Designation; Table 2-2: Model 400E Internal O; Table 2-3: Specifications For Model 400E Izs Generator W/O Reference Feedback Option - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Specifications, Agency Approvals, and
Warranty

Table 2-2: Model 400E Internal O

Maximum Concentration
Minimum Concentration
Initial Accuracy
Stability (7 Days)
Repeatability (7 days)
Response Time
Resolution
Table 2-3: Specifications for Model 400E IZS Generator w/o Reference
Maximum Concentration
Minimum Concentration
Initial Accuracy
Stability (7 Days)
Repeatability (7 days)
Response Time
Resolution

2.2. EPA Equivalency Designation

Advanced Pollution Instrumentation, Inc., Model 400E Ozone Analyzer is designated
as Equivalent Method Number EQOA-0992-087 as defined in 40 CFR Part 53, when
operated under the following conditions:
1. Range: Any range from 100 ppb to 1 ppm.
2. Ambient temperature range of 5 to 40°C.
3. Line voltage range of 105 – 125 and 200 – 240 VAC, 50/60 Hz.
4. With 5-micron PTFE filter element installed in the internal filter assembly.
5. Sample flow of 800 ± 80 cc/min at sea level.
6. Internal or External sample pump.
18
Generator Specifications
3
1.0 PPM
0.050 PPM
+/- 5% of target concentration
1% of reading
1% of reading
< 5 min to 95%
0.5 ppb
Feedback Option
1.0 PPM
0.050 PPM
+/- 10% of target concentration
2% of reading
2% of reading
< 5 min to 95%
0.5 ppb
Model 400E Ozone Analyzer
Instruction Manual
P/N 04316 Rev: B

Advertisement

Table of Contents
loading

Table of Contents