Teledyne 400E Instruction Manual page 91

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
Adjusting the signal zero and span values of the current loop output is done by
raising or lowering the voltage of the respective analog output. This proportionally
raises or lowers the current produced by the current loop option.
Similar to the voltage calibration, the software allows this current adjustment to be
made in 100, 10 or 1 count increments. Since the exact current increment per
voltage count varies from output to output and from instrument to instrument, you
will need to measure the change in the current with a current meter placed in series
with the output circuit (Figure 6-6).
Do not exceed 60 V between current loop outputs and instrument ground.
P/N 04316 Rev: B
See Table 3-2 for
pin assignments of
the Analog Out
connector on the
rear panel.
V OUT +
V OUT -
Analyzer
Figure 6-6:
Setup for Calibrating Current Outputs
mA
IN
OUT
I IN +
I IN -
Recording
Device
NOTE
Operating Instructions
91