REPORT PERIODS IN PROGRESS WHEN INSTRUMENT IS POWERED OFF
If the instrument is powered off in the middle of a REPORT PERIOD, the samples
accumulated so far during that period are lost. Once the instrument is turned back
on, the iDAS restarts taking samples and temporarily them in volatile memory as
part of the REPORT PERIOD currently active at the time of restart. At the end of
this REPORT PERIOD only the sample readings taken since the instrument was
turned back on will be included in any AVG, MIN or MAX calculation. Also, the
STORE NUM. SAMPLES feature will report the number of sample readings taken
since the instrument was restarted.
To define the REPORT PERIOD, follow the instruction shown in section 6.11.2.2
then press:
Use the PREV and NEXT
keys to scroll to the data
channel to be edited.
Set the number of days
between reports (0-366).
Press keys to set hours
between reports in the format :
HH:MM (max: 23:59). This is a
24 hour clock . PM hours are 13
thru 23, midnight is 00:00.
Example 2:15 PM = 14:15
P/N 04316 Rev: B
From the DATA ACQUISITION menu
Edit Data Channel Menu
0) CONC:
ATIMER,
SETUP X.X
PREV NEXT
INS
DEL
NAME:CONC
SETUP X.X
<SET
SET>
EDIT
PRINT
Press SET> key until you reach REPORT PERIOD ...
SETUP X.X
REPORT PERIOD:000:00:05
<SET
SET>
EDIT
PRINT
REPORT PERIODD:DAYS:0
SETUP X.X
0
0
0
SETUP X.X
REPORT PERIODD:TIME:01:01
0
0
0
5
IIf at any time an illegal entry is selected (e.g., days > 366)
the ENTR key will disappear from the display.
2,
4032, R
EDIT
PRNT
EXIT
EXIT
EXIT
ENTR
EXIT
ENTR
EXIT
Operating Instructions
Exits to the main
Data Acquisition
menu.
ENTR accepts the new string and
returns to the previous menu.
EXIT ignores the new string and
returns to the previous menu.
157